Reinhard Schmidt Ep Med Systems

posted October 29, 2009 - 10:56am
Reinhard Schmidt Ep Med Systems

No stimulation channel on screen

* The markers for sites A and B on the back of the stimulator must be connected via a T-connector and a BNC cable to the PMI junction box input marked "PACE DETECT".
* Make sure the setup has PMI- for the negative pin and the correct channel selected for the positive pin. Make sure the analysis is set to STIM. To make it look square, filter between high pass DC and low pass 500Hz.

 

Reinhard Schmidt Ep Med Systems

RF noise is excessive

* The RF generator must be plugged into the EP-WorkMate isolation transformer. The ground pad must not be in close proximity to the surface ECG leads. The RF catheter cannot be used as the reference signal input into the junction box.
* Contact technical support.

Measurements are not displayed in the measurement menu

* Click on the EDIT button and check off the ones you want to add to the menu. Use the Type selection in the measurement window to select the measurement subset most appropriate to the type of study you are doing.

Reinhard Schmidt Termination West Berlin, NJ (October 26, 2004)--EP MedSystems, Inc. (NASDAQ: EPMD), a market leader in electrophysiology products used in cardiac rhythm management, today announced financial results for the three months ended September 30, 2004. Revenues for the quarter were $4,677,000, an increase of 116% from the $2,167,000 reported for the three months ended September 30, 2003, and a 30% increase when compared to the second quarter of 2004. The net loss for the quarter decreased 65%, to $710,000, when compared to the $2,002,000 reported in 2003, and decreased 59% when compared to the $1,748,000 reported for the second quarter of 2004. The basic and diluted loss per share for the three months ended September 30, 2004, was ($0.03) which compared to ($0.11) for the comparable period in 2003. The Company finished the quarter with cash and cash equivalents of $5,783,000.

 

Ep Medsystems Termination

Don't miss out, learn more: Reinhard Schmidt Ep Med Systems

 



Comments

Post new comment

  • Lines and paragraphs break automatically.
  • You can use BBCode tags in the text. URLs will automatically be converted to links.
  • Allowed HTML tags: <p> <br> <b> <a> <em> <strong> <cite> <code> <ul> <ol> <li> <dl> <dt> <dd> <img> <span> <object> <param> <embed> <table> <tr> <td> <div>
  • Web page addresses and e-mail addresses turn into links automatically.

More information about formatting options

Join Xomba Today

Do you like to write? Would you like to make a little extra money on the side? These people do. Join the Xomba community today.
Become a Member